• Home
  • Current congress
  • Public Website
  • My papers
  • root
  • browse
  • IAC-14
  • C2
  • 6
  • paper
  • The laboratory calibration of a novel surface potential monitor at GEO

    Paper number

    IAC-14,C2,6,13,x26447

    Author

    Dr. Chaoyang Zhou, Lanzhou Institute of Physics, China

    Coauthor

    Dr. Shengsheng Yang, Lanzhou Institute of Physics, China

    Coauthor

    Dr. Qin Xiaogang, China

    Year

    2014

    Abstract
    A novel surface potential monitor (SPM) has been developed for a satellite at GEO in China. The SPM consisted of a surface charge detector head, an electrostatic voltmeter and associated control electronics. This novel SPM has been calibrated in a simulated space environment in our laboratory. This paper describes the calibration system which consists of a non-contact voltmeter and an electron beam current monitor. The calibration method is also given, with which the linearity error of SPM is proved less than 1.7% with the measurement range from 1 V to 10 kV and resolution less than 1V in long term.
    Abstract document

    IAC-14,C2,6,13,x26447.brief.pdf

    Manuscript document

    IAC-14,C2,6,13,x26447.pdf (🔒 authorized access only).

    To get the manuscript, please contact IAF Secretariat.