The laboratory calibration of a novel surface potential monitor at GEO
- Paper number
IAC-14,C2,6,13,x26447
- Author
Dr. Chaoyang Zhou, Lanzhou Institute of Physics, China
- Coauthor
Dr. Shengsheng Yang, Lanzhou Institute of Physics, China
- Coauthor
Dr. Qin Xiaogang, China
- Year
2014
- Abstract
A novel surface potential monitor (SPM) has been developed for a satellite at GEO in China. The SPM consisted of a surface charge detector head, an electrostatic voltmeter and associated control electronics. This novel SPM has been calibrated in a simulated space environment in our laboratory. This paper describes the calibration system which consists of a non-contact voltmeter and an electron beam current monitor. The calibration method is also given, with which the linearity error of SPM is proved less than 1.7% with the measurement range from 1 V to 10 kV and resolution less than 1V in long term.